One of the common techniques for converting LaCoste-Romberg survey gravity meters into fedback continuously recording tidal gravity meters is to detect the mass position with a capacitor bridge and apply feedback forces electrostatically. In general, this servosystem is nonlinear and the output voltage has a quadratic as well as a linear dependence on gravity. For the tidal variation in gravity, the quadratic term can be reduced from typically 10% to 0.1% of the linear term with a modification of the standard feedback system. The linearized, fedback gravity meter is precisely calibrated by a two-step procedure. First, the tilt method is used to determine the zero frequency response. The frequency dependence of the transfer function, vital for accurate phase calibration at tidal frequencies, is measured by inserting a random signal in the feedback loop and cross-correlating the random signal with the gravity meter output. The accuracy of the calibration at tidal frequencies is ±0.3% in amplitude and ±02° in phase.