Answers are grounded in this paper and cite their sources.
The Kensey-Nash device is used to fragment gallstones percutaneously. It is inserted into the gallbladder through an 8-F introducer sheath, a cage is expanded to protect the gallbladder, and the impeller is activated. A strong vortex automatically pulls the stones into the basket. The device was tested in 24 pigs in which human gallstones (5-19 mm, bilirubin and cholesterol types) had been implanted and in two control pigs with no gallstones. All pigs except two were killed immediately after the procedure (acute studies); the two were followed up for 36 days and then killed. Fragments (1-3 mm) were similar in size to those seen in current extracorporeal shock wave lithotripsy studies. A complication occurred in one animal, and technical failures occurred in two, causing these three studies to be aborted.
No takes yet. Share an insight, caveat, or question.
Miller et al. (1989) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: