Key result
The BPLab ambulatory blood pressure monitor met all ESH-2001 requirements for accuracy compared to expert BP indices in pregnant women.
Why the study?
Does the BPLab ambulatory blood pressure monitor accurately measure blood pressure in pregnant women compared to expert measurement?
Observational (n=33)
Does the BPLab ambulatory blood pressure monitor accurately measure blood pressure in pregnant women compared to expert measurement?
The BPLab ambulatory blood pressure monitor is validated for use in pregnant women according to ESH-2001 criteria.
Answers are grounded in this paper and cite their sources.
May support BPLab ABPM validation in pregnancy; leaves open need for outcome trials before clinical adoption.
Below are presented the results of BPLab ambulatory blood pressure monitor trials in compliance with International Protocol ESH-2001 method conducted on a group of 33 women with different pregnancy periods. The data on distribution of divergence between expert’s and device’s BP indices testify that the BPLab ambulatory blood pressure monitor meets all the requirements of the said document and may be recommended for wide application for the given category of patients.
No takes yet. Share an insight, caveat, or question.
Бартош et al. (2006) conducted an observational in Pregnancy (n=33). BPLab ambulatory blood pressure monitor vs. Expert's BP indices was evaluated on Distribution of divergence between expert's and device's BP indices. The BPLab ambulatory blood pressure monitor met all ESH-2001 requirements for accuracy compared to expert BP indices in pregnant women.