Higher magnetic field strength in magnetic resonance imaging (MRI) systems offers higher signal-to-noise ratio, contrast, and spatial resolution in magnetic resonance (MR) images. However, the wavelength in ultrahigh fields (7 T and beyond) becomes shorter than the human body at the Larmor frequency with increasing static magnetic field (B0) of the MRI system. At short wavelengths, an interference effect appears, resulting in nonuniformity of the RF magnetic near-field (B1) over the subject and MR images may have spatially anomalous contrast. The B1near-field generated by the transverse electromagnetic RF coil's microstrip line element has a maximum near the center of its length and falls off towards both ends. In this study, a double trapezoidal-shaped microstrip transmission line element is proposed to obtain uniform B1field distribution by gradual impedance variation. Two multi-channel RF head coils with uniform and trapezoidal shape elements were built and tested with a phantom at 7-T MRI scanner for comparison. The simulation and experimental results show stronger and more uniform B1+near-field with the trapezoidal shape.