Why the study?
Case reports detailing the effects of targeted intraoperative radiation therapy (IORT) on patients with pacemakers are rare, and standardized guidelines are lacking.
Does targeted intraoperative radiotherapy cause device malfunction in breast cancer patients with implanted cardiac pacemakers?
Does targeted intraoperative radiotherapy cause device malfunction in breast cancer patients with implanted cardiac pacemakers?
Targeted intraoperative radiotherapy can be safely performed in patients with implanted cardiac pacemakers when safe distances and radiation thresholds are maintained.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for targeted IORT safety with pacemakers; leaves open prospective validation before broader adoption.
Case reports detailing the effects of targeted intraoperative radiation therapy (IORT) on patients with cardiac pacemakers (PMs) are rare. This growing population sub-group requiring IORT and lack of standardized guidelines necessitate more practical published research. An 81-year-old patient with clinical stage II, T1 N0 grade III, triple-negative invasive ductal carcinoma and an implanted single-lead chamber PM (VVIR mode, model: Biotronik, type Effecta SR) received targeted intraoperative radiotherapy at the time of wide local excision and sentinel lymph node biopsy. It presents the shortest distance between the outer diameter of the PM and IORT applicator in literature. Target IORT was performed utilizing an Intrabeam device (50 kV, Carl Zeiss Surgical, Oberkochen, Germany). This case elucidates the successful use of targeted IORT for breast-conserving surgery in a patient with a single ipsilateral chamber VVIR mode PM. No device failure or malfunction was reported for the PM before, during, or after the procedure. These findings support the use of targeted IORT for patients diagnosed with early-stage breast carcinomas who have a PM implanted. However, further research is needed to understand the safety of other methods and devices for IORT patients with cardiac implantable electronic devices.
No takes yet. Share an insight, caveat, or question.
Ramdas et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: