Although the IPTS-68 is meant to be a close approximation to thermodynamic temperatures, its definition makes no provision for continuity in d t /d t 68 at 630.74°C. Any discontinuity in d t /d t 68 at 630.74°C is directly related to the corresponding discontinuity in the first derivative with respect to t 68 of sensor output, so the former can be obtained from measurement of the latter. The magnitude of the discontinuity has not previously been measured precisely; it has an important bearing on elucidating the behaviour of ( t - t 68 ) near 630.74°C. Over the temperature range from about 610°C to 645°C two calibrated platinum resistance thermometers, nine platinum - 10% rhodium/platinum thermocouples, and two other types of thermocouples have been intercompared in an experimental arrangement designed expressly to measure the discontinuity. The result for the relative discontinuity in d t /d t 68 at 630.74°C was found to be 0.51 ± 0.10 percent.