Answers are grounded in this paper and cite their sources.
A recording electron paramagnetic resonance absorption spectrometer of very high sensitivity, operating at a wavelength of 3 cm, is described. It is of the reflection-cavity-in-magic-T-bridge type, using a superheterodyne receiver. Its special features are: (1) a two-cavity klystron oscillator with superior frequency stability and more available power; (2) a synchrodyne system to derive local oscillator power, thus avoiding the difficulty of tracking two independent oscillators; (3) a balanced magic-T mixer which considerably reduces the noise output; (4) a double-input low-noise intermediate frequency amplifier at 30 Mc; and (5) a tunable high-Q sample-carrying cavity with an almost identical cavity to balance it in the magic-T bridge. Arrangements for an oscilloscopic display of the absorption vs magnetic field and a continuous recording of the derivative of absorption vs magnetic field have been incorporated. Sensitivity of the microwave bridge system and of the spectrometer have been tested. About 1.1×10−11 mole of diphenyl picryl hydrazyl dissolved in purified benzene, which corresponded to approximately 1.06×10−14 mole of crystalline sample, has been detected by the spectrometer.
No takes yet. Share an insight, caveat, or question.
Harihar Misra (1958) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: