This is the authors' abstract. We don't add key points for this paper.
Optically pumped magnetometers (OPMs) are a promising magnetoencephalography (MEG) technology for the non-invasive measurement of human electrophysiological signals. Prior work developed biplanar background field-nulling coils necessary for OPM operation, but these were expensive to produce and required tedious error-prone manual winding of >1 km of copper wire. Here, we developed a precise and reproducible manufacturing process by fabricating these coils on two-layer printed circuit boards (PCBs). Building on open-source software (bfieldtools), we developed a pipeline to determine the optimal current loops of 1.5 × 1.5 m2 biplanar nulling coils, connected these loops into a continuous conducting path across PCB layers, and printed them as pairs of 1.5 × 0.75 m2 PCBs, which were soldered and mounted on an aluminum frame. Our coils achieved efficiencies of 1.3–7.1 nT/mA, similar to or higher than previous designs. We reduced the largest background field component from 21 to 2 nT, enabling OPMs in a lightly shielded room to record somatosensory evoked fields (SEFs) comparable to SQUID-MEG. Our coil system is cheaper than commercial alternatives and is available as an open-source package opmcoils, thus enabling more affordable background field nulling for OPM-MEG and realizing its potential as an accessible sensor technology for human neuroscience.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Jas et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: