Key result
The PSF-shape-based beamforming grid combined with 2D cubic interpolation showed the most accurate and stable performance for assessing blood flow and vessel wall dynamics.
A novel PSF-shape-based beamforming strategy with 2D interpolation improves the accuracy of motion estimation in ultrafast ultrasound data.
Answers are grounded in this paper and cite their sources.
Supports optimization of ultrasound beamforming for vascular research; leaves open clinical translation pending human validation.
This paper presents a framework for motion estimation in ultrafast ultrasound data. It describes a novel approach for determining the sampling grid for ultrafast data based on the system’s point-spread-function (PSF). As a consequence, the cross-correlation functions (CCF) used in the speckle tracking (ST) algorithm will have circular-shaped peaks, which can be interpolated using a 2D interpolation method to estimate subsample displacements. Carotid artery wall motion and parabolic blood flow simulations together with rotating disk experiments using a Verasonics Vantage 256 are used for performance evaluation. Zero-degree plane wave data were acquired using an ATL L5-12 (fc = 9 MHz) transducer for a range of pulse repetition frequencies (PRFs), resulting in 0–600 µm inter-frame displacements. The proposed methodology was compared to data beamformed on a conventionally spaced grid, combined with the commonly used 1D parabolic interpolation. The PSF-shape-based beamforming grid combined with 2D cubic interpolation showed the most accurate and stable performance with respect to the full range of inter-frame displacements, both for the assessment of blood flow and vessel wall dynamics. The proposed methodology can be used as a protocolled way to beamform ultrafast data and obtain accurate estimates of tissue motion.
No takes yet. Share an insight, caveat, or question.
Saris et al. (2018) studied this question. PSF-shape-based beamforming grid combined with 2D cubic interpolation vs. Conventionally spaced grid combined with 1D parabolic interpolation was evaluated on Accuracy and stability of motion estimation for blood flow and vessel wall dynamics. The PSF-shape-based beamforming grid combined with 2D cubic interpolation showed the most accurate and stable performance for assessing blood flow and vessel wall dynamics.
Synapse has enriched 10 related papers. Consider them for comparative context: