Key result
Direct aortic valve-in-valve TAVI lowers mean gradient to ~9 mmHg in Leriche syndrome.
Case Report (n=1)
No
Direct aortic TAVI can be successfully performed as a valve-in-valve procedure in patients with severe peripheral arterial disease such as Leriche syndrome.
Answers are grounded in this paper and cite their sources.
May support direct aortic ViV-TAVI when peripheral access is precluded; single-case evidence leaves open generalizability and durability.
Transcatheter aortic valve implantation (TAVI) has been designed to treat elderly patients with severe aortic stenosis considered high-risk surgical candidates. Due to the large device size, even of current generation systems, the trans-femoral approach requires favorable ilio-femoral arterial anatomy; this approach is contraindicated in patients with excessive atherosclerosis, calcifications, or tortuosity of ilio-femoral arteries. We report the case of a 75-year-old female affected by Leriche syndrome who successfully underwent direct aortic (DAo) transcatheter aortic valve implantation with a Medtronic CoreValve Evolut R (Medtronic Inc., Minneapolis, MN, USA) as valve-in-valve implantation to treat a degenerate aortic bioprosthesis.
No takes yet. Share an insight, caveat, or question.
A 2016 study conducted a case report in Degenerate aortic bioprosthesis and Leriche syndrome (n=1). Direct aortic valve-in-valve transcatheter aortic valve implantation with Medtronic CoreValve Evolut R (23 mm) was evaluated on Invasive transvalvular mean gradient. Direct aortic valve-in-valve TAVI with a 23 mm CoreValve Evolut R reduced the trans-prosthesis mean gradient from 100 mmHg to 9 mmHg with no paravalvular leak in a patient with Leriche syndrome.
Synapse has enriched 10 related papers. Consider them for comparative context: