Answers are grounded in this paper and cite their sources.
Bismuth is determined in blood and urine by means of atomic-absorption spectrophotometry following generation of bismuth hydride from a wet-oxidised sample. Recoveries are shown to be satisfactory, and the detection limit using analytical-reagent grade reagents is less than 0.01 µg ml–1.The direct determination of bismuth in urine by means of hydride generation has been investigated, and it is shown that there is severe suppression and poor recovery.
No takes yet. Share an insight, caveat, or question.
R. C. Rooney (1976) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: