Answers are grounded in this paper and cite their sources.
The choice of caesium 137 as a source of γ-ray beams for radiotherapy, and the preparation of suitable sources from the fission products of uranium 235, are discussed in preceding papers (Eastwood, 1960; Wheatley, 1960). The prospect that large sources of this isotope would become readily available stimulated the design of suitable therapeutic equipment. In order that some practical assessment could be made of the value and place of caesium 137 in radiotherapy, a treatment unit was installed at the Downs Branch of the Royal Marsden Hospital. The unit was loaded in October, 1956, with the first of several 137Cs sources produced in the country for tele-therapy apparatus. The source, a cylinder 3·9 cm in length and 2·7 cm in diameter, contained 1,500 c and gave, when first loaded, an output of 23·9 r/minute in air at 40 cm, the source-skin distance used in treatment. A general view of the unit and its main constructional features are shown in Figs. 1 and 2 respectively. It consists essentially of the rotating source holder, the lead shielding and the field defining system. The caesium source (1) is enclosed in a cylindrical platinum container 3·3 cm in diameter and 9·3 cm long which has walls 1·5 mm thick. It is mounted inside a uranium bar (4) 17·5 cm long and of rectangular cross-section 7·5 × 5 cm. This bar can be rotated through 180 deg.
No takes yet. Share an insight, caveat, or question.
Wheatley et al. (1960) studied this question.