Answers are grounded in this paper and cite their sources.
The design and implementation of a scanning near-field optical microscope expressly intended for the imaging of magnetic domains in reflection with subwavelength resolution is described. The microscope employs ∼40 nm silver particles optically excited at the plasmon resonance as the near-field probes. Special attention is given to the Newton ring interferometer used to regulate the separation between the near-field probe and the sample. The results of two tests of the microscope performance are presented. The first is a measurement of the dependence of the near-field magneto-optic Kerr effect on probe/sample spacing for a longitudinally magnetized metallic thin film. The effect decreases rapidly with increasing separation with a length scale comparable to the probe size. The second is a series of images of a domain in a perpendicularly magnetized multilayer metallic film at different probe/sample spacings. The images indicate a persistent effect with increasing separation but diminishing resolution. Fundamental differences between the near-field magneto-optic Kerr effect for perpendicularly and longitudinally magnetized samples is discussed. An estimate is made as to the fundamental limit of resolution with this particular form of near-field microscopy.
No takes yet. Share an insight, caveat, or question.
Silva et al. (1996) studied this question.