Key result
TOF-Cuff placed on the lower leg showed no significant difference in mean time to a TOF ratio of 0% compared to TOF-Scan on the forearm (p=0.1356), but demonstrated large limits of agreement and a 25% measurement failure rate.
Why the study?
The study was conducted to compare two devices for neuromuscular monitoring during anesthetic induction: TOF-Cuff installed on the lower leg versus conventional TOF-Scan over the ulnar nerve.
Does TOF-Cuff on the lower leg provide comparable neuromuscular monitoring to TOF-Scan on the wrist in adult patients undergoing shoulder surgery?
Observational (n=20)
No
Does TOF-Cuff on the lower leg provide comparable neuromuscular monitoring to TOF-Scan on the wrist in adult patients undergoing shoulder surgery?
TOF-Cuff placed on the lower leg showed large limits of agreement and a high failure rate compared to standard TOF-Scan on the wrist for neuromuscular monitoring during anesthetic induction.
Lower-leg TOF-Cuff shows poor agreement and high failure versus forearm monitoring; leaves open its reliability for routine use.
Purpose The aim of this study was to compare two devices for neuromuscular monitoring during anesthetic induction. TOF-Cuff® was installed on the lower leg stimulating the tibial nerve, while the more conventional TOF-Scan® was installed over the ulnar nerve at the wrist. Methods Twenty adult patients were enrolled in this prospective, controlled study. Train-of-four (TOF) was recorded every 15 s until TOF ratio of 0%. Mean arterial blood pressure (MAP) was assessed with TOF-Cuff® and with standard anesthesia monitoring from the brachial artery. MAP was measured before and after anesthetic induction. Time to TOF ratio = 0% was compared with one-sample t test and Bland–Altman plots. Results Patients received 0.53 ± 0.09 mg atracurium per kg body weight intravenously. Mean time to TOF ratio = 0% was 150.8 s (± 43.7) for TOF-Scan®, and 174.4 s (± 42.7) for TOF-Cuff® (p = 0.1356). Bias was − 15.9 (95% confidence interval − 37.5 to 5.6) with 95% limits of agreement of − 95.2 to 63.3. Twenty-five percent of the patients had a technical issue with a TOF-Cuff® measurement. For MAP, mean difference was 1.4 (95% confidence interval − 2.4 to 5.2) with 95% limits of agreement of − 22.7 to 25.5. Conclusion The time from administration of a common dose of atracurium to a TOF ratio of 0% assessed with TOF-Cuff® stimulating the tibial nerve compared to TOF-Scan® stimulating the ulnar nerve showed large limits of agreement in Bland–Altman analysis. There was a high failure rate with TOF-Cuff® measurements on the lower leg.
No takes yet. Share an insight, caveat, or question.
Dullenkopf et al. (2019) conducted an observational in Shoulder surgery requiring general anesthesia (n=20). TOF-Cuff® on the lower leg vs. TOF-Scan® on the forearm was evaluated on Time from administration of neuromuscular blocking agents to a TOF ratio of 0% (MD -15.9, 95% CI -37.5 to 5.6, p=0.1356). TOF-Cuff placed on the lower leg showed no significant difference in mean time to a TOF ratio of 0% compared to TOF-Scan on the forearm (p=0.1356), but demonstrated large limits of agreement and a 25% measurement failure rate.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: