We report a terahertz time-domain spectrometer with more than 6 THz spectral coverage and 1 GHz resolution based on high-speed asynchronous optical sampling. It operates at 2 kHz scan rate without mechanical delay stage. The frequency error of the system at 60 s acquisition time is determined by comparing a measured water vapor absorption spectrum to data reported in the HITRAN database. The mean error of 87 evaluated absorption lines is 142 MHz.
No takes yet. Share an insight, caveat, or question.
Klatt et al. (2009) studied this question.