A Schwarzschild-type x-ray microscope has been designed, constructed, and tested. Ni/C multilayers were used as the x-ray mirrors, with a thickness (2d) of 7 nm and 30 layer pairs. The microscope has attained a spatial resolution of 0.5 microm at a magnification of 15. By using bright laser-produced plasmas as an x-ray source [R. Kodama, K. Okada, N. Ikeda, M. Mineo, K. A. Tanaka, T. Mochizuki, and C. Yamanaka, J. Appl. Phys. 59, 3050 (1986)], images could be recorded during the 400-psec laser pulse.