Key result
Non-invasive intrapartum fetal ECG successfully acquired fetal ECG signals in 80% (12/15) of patients during labour.
Why the study?
Can a non-invasive device using 12 abdominal electrodes successfully acquire detailed intrapartum fetal ECG signals and uterine contractions?
Observational (n=15)
No
Can a non-invasive device using 12 abdominal electrodes successfully acquire detailed intrapartum fetal ECG signals and uterine contractions?
A non-invasive abdominal electrode system can successfully acquire detailed intrapartum fetal ECG waveforms and heart rate variability in the majority of patients, potentially avoiding invasive fetal scalp electrodes.
Answers are grounded in this paper and cite their sources.
Demonstrates technical feasibility of non-invasive fetal ECG; leaves open clinical utility and need for larger validation trials.
OBJECTIVES: To obtain fetal heart rate, detailed fetal electrocardiography (fECG) signals and uterine contractions during labour using a single device. DESIGN: Prospective observational study. SETTING: Delivery suite at a tertiary referral hospital, London, UK. POPULATION: Fifteen patients at median gestation of 39 weeks (range 24-41) were recruited at median cervical dilatation of 4.0 cm (range 0-10) of whom 8/15 (53%) had intact amniotic membranes. METHODS: Using 12 abdominally sited electrodes, we recorded the composite abdominal signal in pregnancies intrapartum. The recorded data were analysed off-line using a blind signal separation technique. MAIN OUTCOME MEASURES: Success of signal separation and fECG time intervals. RESULTS: Successful fECG signal acquisition was achieved in 12/15 (80%) patients and an averaged fECG waveform acquired. In these patients, P and QRS waves were seen in all cases, and T waves in 11/12 (92%). True beat-to-beat heart rate (HR) was displayed and measures of its variability obtained. The mother's ECG and uterine electrical activity, shown to match tocographically recorded uterine contractions, were also separated and displayed. Failure to acquire fECG in three cases was attributed to excessive abdominal muscular activity and electrical interference. CONCLUSIONS: This study demonstrates a non-invasive technique that displays detailed intrapartum fECG waveforms, HR variability, maternal ECG and uterine contractions simultaneously, all in a single device and which avoids the potential risks of invasive monitoring with a fetal scalp electrode.
No takes yet. Share an insight, caveat, or question.
Taylor et al. (2005) conducted an observational in Labour (intrapartum) (n=15). Non-invasive intrapartum fetal ECG was evaluated on Success of signal separation and fECG time intervals. Non-invasive intrapartum fetal ECG successfully acquired fetal ECG signals in 80% (12/15) of patients during labour.