Key result
In an in vitro porcine model, effective orifice area decreased significantly with SCA, IR, and ER annuloplasty compared to baseline (P<0.001), with higher transvalvular pressure drops in ER and IR vs SCA (P=0.007).
Why the study?
Do different annuloplasty techniques (SCA, IR, ER) differentially affect aortic valve function and root morphology in an in vitro porcine model?
Do different annuloplasty techniques (SCA, IR, ER) differentially affect aortic valve function and root morphology in an in vitro porcine model?
p-value: p=<0.001
In an in vitro porcine model, external and internal ring annuloplasty techniques demonstrated greater potential to reduce ventriculo-aortic junction diameter compared to subcommissural annuloplasty, but induced different root remodeling patterns.
Should not alter annuloplasty selection in patients; leaves open whether SCA preserves hemodynamics better in vivo.
OBJECTIVES: In aortic valve repair, reduction and stabilization of the ventriculo-aortic junction (VAJ) is generally recommended. In this in vitro study, we compare three techniques of annuloplasty: the subcommissural annuloplasty (SCA), the internal ring (IR) and the external ring (ER) annuloplasty. METHODS: Ten fresh porcine aortic valve preparations were tested in a pulsatile mock loop. Each sample was tested untreated (baseline: B). The annuloplasty techniques were then performed successively in each sample. Each technique was tested, then removed and the following technique performed. SCA was applied at 50% of interleaflet triangle height; the ER and IR were applied with a moderate reduction (15-20%) of the VAJ. Hydrodynamic, video and echographic parameters were collected. Flow rate and arterial pressure were maintained consistently between groups. RESULTS: Effective orifice area decreased significantly with each annuloplasty technique compared with baseline (P < 0.001). Mean transvalvular pressure drop was significantly higher in the ER and IR vs SCA (P = 0.007). Annuloplasty reduced valve opening and closing time in comparison to baseline. Echocardiography confirmed that the VAJ experienced a greater reduction with the ER and IR vs SCA. A narrowing of the lower third of the sinuses of Valsalva was observed after the ER, and subvalvular narrowing was observed after the IR. Valve coaptation increased with all annuloplasty techniques. CONCLUSIONS: The three annuloplasty techniques examined demonstrated differential effects on aortic valve function and root morphology. The ER and IR have greater potential to reduce VAJ diameter in comparison to SCA. The IR induced a subvalvular remodelling of the VAJ, whereas the ER induced a paravalvular remodelling.
No takes yet. Share an insight, caveat, or question.
Kerchove et al. (2012) studied Aortic valve repair (in vitro model) (n=10). Subcommissural annuloplasty (SCA), internal ring (IR), and external ring (ER) annuloplasty vs. Untreated baseline and each other was evaluated on Effective orifice area and mean transvalvular pressure drop (p=<0.001). In an in vitro porcine model, effective orifice area decreased significantly with SCA, IR, and ER annuloplasty compared to baseline (P<0.001), with higher transvalvular pressure drops in ER and IR vs SCA (P=0.007).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: