Answers are grounded in this paper and cite their sources.
We demonstrate the feasibility of swept-source visible-light optical coherence tomography (SS-vis-OCT). We used a fanout periodically poled lithium niobate (PPLN) crystal for second-harmonic generation (SHG) to convert a commercial near-infrared swept-source (NIR-SS) laser into a visible-light SS laser. To enhance SHG efficiency, we amplified the NIR-SS output with booster optical amplifiers (BOAs) and generated up to 580 µW of power. The SS-vis-OCT achieved a maximum axial resolution of 7.3 µm and an imaging depth of 5 mm in air, corresponding to 5.4 µm and 3.7 mm in tissue (n = 1.35). Compared with spectral-domain vis-OCT, SS-vis-OCT provides a 2.2-fold increased imaging depth and a 2.8-fold improved roll-off. Additionally, we validated SS-vis-OCT performance using a 3D-printed pyramid phantom, with its feature measurements cross-validated by scanning electron microscopy (SEM).
No takes yet. Share an insight, caveat, or question.
Fan et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: