Key result
Ethanol was superior to DMSO for assisting lead disconnection from pacemaker or ICD headers, yielding successful disconnection at the primary attempt in of cases.
Why the study?
Does ethanol improve successful lead disconnection compared to DMSO in patients with stuck pacemaker or ICD leads during battery replacement?
RCT (n=30)
randomly assigned
Does ethanol improve successful lead disconnection compared to DMSO in patients with stuck pacemaker or ICD leads during battery replacement?
Ethanol is a simple, inexpensive, and highly effective solvent for facilitating lead disconnection from ICD or pacemaker headers when conventional removal fails.
Answers are grounded in this paper and cite their sources.
May aid challenging lead disconnections in revisions; hypothesis-generating pilot data requiring larger randomized confirmation.
AIMS: Battery exchange in pacemaker (PM) or implantable cardioverter defibrillator (ICD) devices may be occasionally problematic because of difficulties in lead disconnection procedures and risk of injuring the fragile leads. This pilot study compares ethanol and dimethyl sulfoxide (DMSO) as solvents to assist removal of leads from PM or ICD device headers in cases of stuck leads or difficulties in untightening device header screws. METHODS AND RESULTS: Of the total number (527) of our patients requiring battery replacement due to end-of-life (EOL) warnings, conventional exchange was not possible in 34 (6.5%) due to embedding of the lead within blood-derived material. Of these, 30 (17 with PM, 13 with ICD) consented to the study and were randomly assigned to a primary attempt at lead disconnection by ethanol (n = 17) or by DMSO (n = 13). If the primary attempt failed, a secondary attempt at lead disconnection was undertaken using the alternate solvent. Ethanol was a superior solvent compared with DMSO, yielding successful disconnection at primary attempt in 88.2% (15/17) vs. 23.1% (3/13) of cases. In 8 patients in whom the primary DMSO-attempted disconnection failed, a secondary attempt with ethanol yielded success in 6 (75%) cases. Use of either ethanol or DMSO in lead disconnection was not associated with any adverse events or effects. CONCLUSION: Ethanol has utility as a simple and inexpensive modality for lead disconnection from ICD or PM headers.
No takes yet. Share an insight, caveat, or question.
Erné et al. (2016) conducted an RCT in Pacemaker or ICD battery replacement with stuck leads (n=30). Ethanol vs. Dimethyl sulfoxide (DMSO) was evaluated on Successful lead disconnection at primary attempt. Ethanol was superior to DMSO for assisting lead disconnection from pacemaker or ICD headers, yielding successful disconnection at the primary attempt in of cases.