Answers are grounded in this paper and cite their sources.
A sorter that is activated by the resistance pulses of a Coulter orifice (volume detector) has been developed. The problem arising from the distortion of the volume signals produced by the electric droplet forming and charging voltages are avoided by special shielding and grounding of the volume detector. Experiments with beads and leukocytes are described. The maximal processing rate at this time is 2000 cells/sec, the purity of sorted fractions is better than 97%, and the viability of sorted cells is better than 80%.
No takes yet. Share an insight, caveat, or question.
Menke et al. (1977) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: