Why the study?
Rapid growth in infants and toddlers creates challenges for aortic valve substitutes, requiring evaluation of hydrodynamic performance and response to annular expansion.
Does a handmade bioprosthetic individualized aortic valve prosthesis improve hydrodynamic performance and accommodate simulated growth compared to a commercially available pulmonary conduit and polyurethane valve in an in vitro model?
Does a handmade bioprosthetic individualized aortic valve prosthesis improve hydrodynamic performance and accommodate simulated growth compared to a commercially available pulmonary conduit and polyurethane valve in an in vitro model?
A novel handmade bioprosthetic aortic valve demonstrated promising in vitro hydrodynamic performance and the ability to accommodate simulated growth for pediatric applications.
Bench performance of handmade pediatric aortic valve; leaves open in vivo durability and safety before any clinical consideration.
This study evaluated the hydrodynamic performance of a handmade bioprosthetic aortic heart valve substitute intended for infants and toddlers. This valve was prepared using neutralized animal tissue and consequently tested in a pre-established pulse duplicator. Valve kinematics were characterized by an in-house developed image analysis algorithm. Analysis results provide a full map of valve's response including mean pressure differential, regurgitation, as well as orifice area during normotensive, hypotensive, and hypertensive conditions. The objective valve prosthesis was also compared with a commercially available pulmonary conduit along with an in-house built polyurethane valve. As growth is very rapid on target group's age, the investigation was further expanded to evaluate valve's response upon expansion (as in growth) of the aortic annulus. The latter was achieved by a novel scheme that uses vacuum conditions on a flexible silicone mock annulus, providing a diameter growth with highly biomimetic characteristics. The results demonstrated that the objective valve had good responsiveness, following norm's requirements in all tested conditions, with marginal deviation on regurgitation, while satisfactory responsiveness was evaluated in expanded (growth) state, too, representing 2 years of aging. Nevertheless, the alternative tested valves struggled to reach the requirements, while the potential on accommodation of growth appears limited particularly for rigid, stented valves.