Answers are grounded in this paper and cite their sources.
A flow injection immunoassay was performed using a column packed with reversed-phase sorbents to effect separation of the immunoreacted species by entrapping free analyte and allowing antibody-conjugated analyte to pass unretained. Fluorescein-labeled analyte was measured in a competitive assay for the anticonvulsant drug phenytoin. The simplicity of the assay was the greatest advantage of the technique, which allowed for measurement of phenytoin in a 2-min assay time. The reliable detection limit for the assay was 5 nmol L(-)(1) of phenytoin in serum. The columns were regenerated with periodic injections of ethanol solutions to remove the entrapped analyte and prepare the column for subsequent analyses.
No takes yet. Share an insight, caveat, or question.
Locascio-Brown et al. (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: