Answers are grounded in this paper and cite their sources.
Eight platinum resistance thermometers satisfying the requirements of the International Temperature Scale were calibrated on the International scale and intercompared from -190 to 445 C. The o's of the Callendar-Van Dusen equations for seven of these thermometers ranged from 1.49375 to 1.49862. The {) for the eighth'jthermometer,IE, was 1.51155. After minimizing the effects of relative calibration errors, the maximum difference between the readin~s of the group of seven thermometers was 7 millidegrees between -190 and 0 C and 1.3 millidegrees between 0 and 100 C. Maximum differences from the mean were 4.0 and 0'8 millidegree, respectively. For thermometer E, deviations from the mean of the other seven were 15 millidegrees at -110 C and 3 millidegrees at + 50 C. Tables of differences between readings of platinum resistan ce thermometers arising from assumed calibration errors at the fixed points were calculated and are included.
No takes yet. Share an insight, caveat, or question.
Höge et al. (1942) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: