Answers are grounded in this paper and cite their sources.
A low-temperature calorimetric technique for heat capacity measurement of small samples is described. The sample is connected by thermal resistance to a thermal sink at a chosen temperature. The precooling of the sample to this starting temperature is simple and fast. The sample temperature rises exponentially in time after electrical heating power is applied to it maintaining the sample in near thermal equilibrium with its surroundings. An analogous electrical circuit of such microcalorimeter is introduced and analysed to establish conditions for its realization and to determine the possibilities for the interpretation of experimental temperature-time curves. More than one sample can be maintained in the cryostat. It enables comparison to be made between the results obtained with an unknown sample and those obtained with standard samples of known heat capacities used as references.
No takes yet. Share an insight, caveat, or question.
Durek et al. (1972) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: