Answers are grounded in this paper and cite their sources.
We report properties of single-wire proportional chambers and multiple-wire ionization chambers filled with liquid xenon. Proportional multiplication is seen for anode fields > 10⁶ V/cm (anode-cathode voltage >1 kV for a 3.5-{μ}m-diam anode wire). The efficiency for detecting ionizing radiation is {}100%. The time resolution of the chambers is ±{}10^-7 sec. We attained a spatial resolution of ±{}15 {μ}m with a multiwire ionization chamber, considerably better than that of any other real-time particle detector.
No takes yet. Share an insight, caveat, or question.
Muller et al. (1971) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: