A viscometer formed of a tungsten wire of radius about 0.2 mm has been evaluated with measurements of the viscosity of certified reference material for viscosity N350 at viscosities up to 1.581 Pa·s, at temperatures between (289.1 and 420.7) K and a pressure of 0.1 MPa. The results differed from the cited values by less than ± 10 %.
No takes yet. Share an insight, caveat, or question.
Etchart et al. (2007) studied this question.