Answers are grounded in this paper and cite their sources.
Ring core-biotinylated testosterone tracers were synthesized with bridges of three different lengths connecting the biotin moiety to the steroid core (7α-Cn-Bio-T, n = 3, 6, or 11). Together with a position 7-specific polyclonal anti-testosterone antibody, we used the 7α-C11-Bio-T tracer to develop a novel, labeled-hapten competitive immunoassay for total testosterone in serum. (The C3 and C6 tracers proved to be not suitable for analogous immunoassays.) Enhanced chemiluminescence signal was generated by use of a second immobilized antibody and a streptavidin–horseradish peroxidase conjugate. The measuring range of the assay is 0.2–20.0 nmol/L, linearity of serial dilutions can be demonstrated, the lower detection limit is 0.125 nmol/L, and the interassay imprecisions are 13–16%. Accuracy determinations in mass spectrometry-controlled reference specimens showed a mean recovery of 95%. In addition, the assay shows low cross-reactivities, demonstrating the favorable specificity of the combination of a “nearly native” tracer with a position analog antibody. The optimized steric structure and the long spacer arm of the biotinylated testosterone tracer make this chemiluminescence assay well-suited for measuring total testosterone concentration in serum.
No takes yet. Share an insight, caveat, or question.
Luppa et al. (1997) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: