Answers are grounded in this paper and cite their sources.
A new enzymatic amperometric sensor for hydrogen peroxide based on a modified carbon paste electrode is described. Horseradish peroxidase and ferrocene mediator were immobilized by dissolution or dispersion in a graphite–paraffin oil paste. To prevent losses of immobilized chemicals under operational conditions, the electrode surface was coated with a Nafion membrane. The device exhibits a relatively fast response time (42 s for 95% of the steady‐state response) and yields currents linearly related to the hydrogen peroxide concentration over the 10−7 to 10−5 mol/L range in both static and flow injection operations. The slope of the calibration plot remained constant for more than 20 days. Flow injection assays of hydrogen peroxide at a sampling rate of about 70 injections per hour, with a relative standard deviation of 1.4%, are possible. An easy and low cost preparation procedure is described.
No takes yet. Share an insight, caveat, or question.
Sánchez et al. (1991) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: