We report a transportable one-dimensional optical lattice clock based on 87Sr at the National Time Service Center. The transportable apparatus consists of a compact vacuum system and compact optical subsystems. The vacuum system with a size of 90 cm× 20 cm× 42 cm and the beam distributors are assembled on a double-layer optical breadboard. The modularized optical subsystems are integrated on independent optical breadboards. By using a 230 ms clock laser pulse, spin-polarized spectroscopy with a linewidth of 4.8 Hz is obtained which is close to the 3.9 Hz Fourier-limit linewidth. The time interleaved self-comparison frequency instability is determined to be 6.3 × 10–17 at an averaging time of 2000 s.