Answers are grounded in this paper and cite their sources.
Tolerance analysis shows that an efficient M(2) measurement plan is a first cut (beam diameter measurement) at 0.5-2.0 Rayleigh ranges to one side of the waist, which is matched by interpolation between second and third cuts to the opposite side. The waist is measured by a fourth cut halfway between the matched diameters, yielding an easy two-parameter curve fit for M(2).
No takes yet. Share an insight, caveat, or question.
Thomas F. Johnston (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: