Answers are grounded in this paper and cite their sources.
Terahertz-wave transmission loss measurement systems at 330–500 GHz using single- and dual-channel superheterodyne receivers are described. It is shown that the phase noise effect in a single-channel system causes significant signal level fluctuation in the lock-in amplifier. The dual-channel system can eliminate the phase noise effect in synchronous detection and achieve a room-temperature noise floor of −170 dBm. The dynamic range of the dual-channel system is estimated to be 140–152 dB. The measurement uncertainty of the system is analyzed. A 0–100 dB WR-2.2 waveguide attenuator can be measured with expanded uncertainties of 0.078–0.24 dB at 480 GHz. The dual-channel system can serve as a national measurement standard for waveguide transmission measurement at 330–500 GHz and provide traceability for measurements made using commercial instruments.
No takes yet. Share an insight, caveat, or question.
Thomas Y. Wu (2012) studied this question.