Purpose: The accuracy and agreement of intraocular pressure (IOP) measurements using a noncontact self-measuring tonometer, TONO-i, compared with the Goldmann applanation tonometer (GAT).Methods: A total of 35 healthy participants and 42 primary open angle or normal tension glaucoma patients were enrolled in this study. TONO-i and GAT measurements were obtained and the mean difference in IOPs was analyzed using a paired t-test. IOP agreement was calculated using the two-way random effects model. Bland-Altman analysis and simple linear regression were used to estimate the disagreement between IOP measurements.Results: The mean IOPs measured using the TONO-i and GAT were 15.9 ± 3.3 mmHg and 16.4 ± 5.1 mmHg, respectively. In glaucoma patients, the mean values measured using the TONO-i (16.7 ± 4.1 mmHg) were lower than those measured using the GAT (17.9 ± 6.1 mmHg; p < 0.001). However, there was no significant difference among the healthy participants (p = 0.273). The intraclass coefficients (95% confidence interval) for all participants, normal participants, and glaucoma patients were 0.847 (0.702-0.920), 0.882 (0.804-0.941), and 0.828 (0.742-0.887), respectively, indicating good agreement between the IOP measurements. The mean difference and limits of agreement were not statistically significant. As IOP increased, the absolute difference between the GAT and TONO-i measurements also increased.Conclusions: The TONO-i, a novel tonometer, may be used reliably for IOP self-measurement and detecting IOP elevations. However, caution must be exercised because of the possibility of underestimating high IOPs.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: