An easily constructed calorimeter is described which can determine specific heats on an absolute basis with an estimated accuracy of 2%, with a sample temperature rise of less than 2°F. Sheets of sample material are rolled with a metal foil heating element into a cylinder resembling a jelly roll. A thin foil thermocouple is inserted at least 0.25 in. from the roll surface. It is established that there is a brief period after heating is begun when the material near the thermocouple is heated adiabatically; from the recorded temperature rise rate and other readily available data, the sample specific heat is computed. The sample assembly is readily placed in a bath or an oven to obtain values at temperatures other than room temperatures. Measurements are reported for polyethylene terephthalate, polyvinyl chloride, and polyethylene, the latter two from liquid nitrogen temperature to decomposition temperatures. It is suggested that the same principles can be used with modified techniques to measure specific heats of conductors, rigid sheets, or liquids.