I.Collaborative assays to determine the amino acid content of a standard mixture of amino acids and of two test proteins, gelatin and freeze-dried cod muscle, were carried out to assess the accuracy of the ion-exchange column chromatographic technique as operated in different laboratories. 2.Nine laboratories took part, four using single-column automatic amino acid analysers, three using two-column automatic amino acid analysers and two using the manual procedure. 3.In the assay of the amino acid mixture the percentage mean absolute deviations in laboratories using single-column and two-column automatic amino acid analysers, and in one laboratory using the manual method, were 4.8, 2.2 and 6.8 respectively. 4.In the assay of acid-hydrolysates of cod muscle the percentage absolute mean deviations from the mean value in laboratories using two-column automatic amino acid analysers were 7.2, 2.9 and 9'4 for single-column and two-column automatic amino acid analysers and the manual method, respectively.5. The corresponding values in the assay of hydrolysates of gelatin were 7.6, 3.7 and 6.1, respectively.6. Possible explanations are put forward for the different degrees of precision achieved by the methods used.