Answers are grounded in this paper and cite their sources.
The condition of the surface of a field emission cathode is not constant with time, but varies as a result of contamination by residual gases and of ion bombardment. This causes emission fluctuations, and consequent variation of the current in the electron probe. Stabilization of the probe current can be achieved by using an electrode following the electron gun to collect a probe current monitor signal, which is used to determine the voltage applied to an emission control electrode in the gun. This enables satisfactory micrographs to be produced under cathode conditions that would otherwise cause substantial emission fluctuations and unacceptable intensity variation across the micrograph. For such a system, requirements are determined for the collector electrode size and the stabilizer system bandwidth. Stabilization has some optical consequences which degrade the system performance, and to avoid these it is sometimes preferable to use the monitor signal to control the video channel gain, to compensate for the probe current fluctuations.
No takes yet. Share an insight, caveat, or question.
J. R. A. Cleaver (1975) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: