Answers are grounded in this paper and cite their sources.
In neonates, extracorporeal membrane oxygenation (ECMO) is increasingly used for circulatory support, e.g., after cardiac surgery. For training purposes and for research, animal experiments are usually required, complicated by increasing social issues, high costs, and limited reproducibility. Therefore, we designed a mechanical neonatal mock circulation (NMC) model enabling pulsatility and various hemodynamic conditions commonly occurring in neonates. Connected to a flow and pressure reading interface, a computer assisted data management system was installed. A nonocclusive roller pump combined with stiff and elastic tubing segments (for aortic pressure regulation and venous capacity) as well as constant and variable resistance (and optionally a patent duct) are essential features of the NMC system. To show the investigational potential, we studied the influence of venoarterial and venovenous ECMO on the NMC performance during normal circulation, hypovolemia, high arterial resistance, the combination of both, and in low cardiac output. By assessing the significant effects of ECMO on the circulatory function of the NMC, its feasibility and investigational properties could be demonstrated.
No takes yet. Share an insight, caveat, or question.
Trittenwein et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: