Answers are grounded in this paper and cite their sources.
This study reports on an improved method for acetaldehyde (ACH) determination in blood by high-performance liquid chromatography (HPLC). In the case of HPLC analysis, ACH is generally converted to derivatives for ultraviolet detection (for example 2,4-dinitrophenylhydrazine [DNPH] derivative). Nevertheless, elevation of the background during protein precipitation, hydrazone synthesis, or both frequently results in a serious loss of accuracy and precision of the analysis. The method in this study is developed to minimize the increase in nonspecific ACH-DNPH with a view to optimize mainly the synthetic condition of ACH-DNPH. The background is decreased dramatically by gentle deproteination, optimization of the DNPH amount and reaction pH, and reversed-phase solid extraction for the elimination of excess DNPH reagent. The standard curves show good linearity between 0 and 100 microM and minimal background is observed, indicating that the method is useful for monitoring the ACH concentration in blood.
No takes yet. Share an insight, caveat, or question.
Kozutsumi et al. (2002) studied this question.