This is the authors' abstract. We don't add key points for this paper.
A T-format, photon-counting polarization photometer was constructed to facilitate precise polarization measurements at low signal-to-noise ratios. The instrument optics and electronics are described. Several examples of the instrument's performance are given, including measurements on picomolar fluorescein solutions at low excitation resolution and the excitation polarization spectrum of indole (2 x 10(-4)M) at a resolution of 5 A.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Jameson et al. (1978) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: