This is the authors' abstract. We don't add key points for this paper.
Pedicle screw stability is poor in osteopenic vertebrae attributable, in part, to low screw-bone interface strength. The current authors examined cement augmentation using a low curing temperature hydroxyapatite and bis-phenol-A glycidol methacrylate-based composite resin. This cement may stiffen the screw-bone interface and reduce the harmful effects associated with polymethylmethacrylate regarding temperature and toxic monomer. Thirty-five lumbar vertebrae from human cadavers were instrumented with pedicle screws, with one pedicle previously injected with cement and the other as the control. Caudocephalad toggling of ± 1 mm for 1600 cycles was applied to the pedicle screws, and the resulting forces supported by the implant-bone interface were captured by a load cell. A curve was constructed from the peak caudal load for each cycle and three mechanical measures parameterized this curve: (1) initial load; (2) rate of load decay during the first 400 cycles; and (3) final load. The initial load increased by 16% as a result of cement augmentation, the final load increased by 65%, and the rate of load decay decreased by 59%. Cement augmentation of pedicle screws increased the stiffness and stability of the screw-bone interface.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Turner et al. (2003) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: