Key result
Maximal carotid intima-media thickness measured by a wireless handheld ultrasound device showed excellent agreement (ICC=0.960) with a standard ultrasound device.
Why the study?
Routine CIMT screening is criticized for low cost-effectiveness in asymptomatic patients, prompting evaluation of an economic wireless handheld ultrasound device for CIMT measurement in symptomatic patients.
Does wireless handheld ultrasound measurement of CIMT provide comparable accuracy to standard ultrasound and improve prediction of coronary artery disease in symptomatic patients?
Cross-Sectional (n=100)
Single-blind
No
Does wireless handheld ultrasound measurement of CIMT provide comparable accuracy to standard ultrasound and improve prediction of coronary artery disease in symptomatic patients?
Wireless handheld ultrasound provides a feasible, reproducible, and potentially cost-effective alternative to standard ultrasound for measuring CIMT to aid in predicting coronary artery disease in symptomatic patients.
Answers are grounded in this paper and cite their sources.
Supports portable carotid IMT assessment in limited settings; leaves open larger validation trials before clinical adoption.
PURPOSE: Routine screening for carotid intima-media thickness (CIMT) and cardiovascular (CV) disease in asymptomatic patients has been criticized for the high costs and large number of patients required for detecting one patient with coronary artery disease (CAD). In order to overcome the low cost-effectiveness thereof, we investigated the feasibility of an economic wireless handheld ultrasound (WHUS) device for CIMT measurement in symptomatic patients. MATERIALS AND METHODS: A total of 100 consecutive patients with cardiac symptoms were enrolled. CIMT was measured in all patients. Coronary angiography was performed in 75 patients indicated for the exam. RESULTS: The mean of maximal CIMT measured from left/right common carotid artery and bulb (max-CIMT) by the WHUS device showed excellent agreement [intraclass correlation coefficient (ICC)=0.960] with a standard ultrasound device and great interobserver repeatability (ICC>0.9 between all observers). Receiver operating characteristic curve analysis showed that the predictive power for CAD was improved when max-CIMT and plaque information (plaque≥2) was added [area under the curve (AUC): 0.838] to the traditional clinical CV risk factors (AUC: 0.769). The cutoff values for CAD prediction with the standard device and the WHUS device were 1.05 mm (AUC: 0.807, sensitivity: 0.78, specificity: 0.53) and 1.10 mm (AUC: 0.725, sensitivity: 0.98, specificity: 0.27), respectively. CONCLUSION: max-CIMT measured by a WHUS device showed excellent agreement and repeatability, compared with standard ultrasound. Combined max-CIMT and plaque information added predictive power to the traditional clinical CV risk factors in detecting high-risk CAD patients.
No takes yet. Share an insight, caveat, or question.
Jang et al. (2020) conducted a cross-sectional in Cardiac symptoms (n=100). Wireless handheld ultrasound (WHUS) device vs. Standard console ultrasound device was evaluated on Maximal carotid intima-media thickness (max-CIMT) measurement agreement (MD -0.158 mm, p=<0.050). Maximal carotid intima-media thickness measured by a wireless handheld ultrasound device showed excellent agreement (ICC=0.960) with a standard ultrasound device.
Synapse has enriched 10 related papers. Consider them for comparative context: