Why the study?
The study aimed to estimate the feasibility and accuracy of carotid intima-media thickness and hemodynamic parameters measurement using a handheld ultrasound device.
Does a handheld ultrasound device provide accurate measurements of carotid artery structure and hemodynamics compared to conventional ultrasound in routine volunteers?
Does a handheld ultrasound device provide accurate measurements of carotid artery structure and hemodynamics compared to conventional ultrasound in routine volunteers?
Handheld ultrasound devices provide moderate to excellent agreement with conventional ultrasound for basic assessment of carotid artery structure and hemodynamics, supporting their use as a diagnostic tool.
Answers are grounded in this paper and cite their sources.
May support handheld carotid assessment in volunteers; leaves open validation in patients and outcome studies.
Objective: The study aimed to estimate the feasibility and accuracy of carotid intima-media thickness (CIMT) and hemodynamic parameters measurement in a handheld ultrasound device. Methods: Utilizing an ex vivo pig carotid artery sample, CIMT was measured with a handheld ultrasound and compared with histopathology. Then we performed a carotid ultrasound on 25 volunteers using a handheld ultrasound device and a conventional ultrasound system. After a week, these volunteers were scanned again by the same observer. Assessments of the max IMT, mean IMT and hemodynamic parameters (PSV, EDV, PI, RI, S/D, ACCEL, AT, TAMEAN) were compared. Intraclass correlation coefficient (ICC) was used to assess inter-device agreement. Repeatability and correlation of mean IMT were analyzed by Bland–Altman Plots and linear correlation analysis. Results: The mean IMT measured from the common carotid artery by handheld ultrasound showed good agreement (ICC=0.79) with conventional ultrasound. Furthermore, we obtained good repeatability and a consistent trend in the mean value of IMT before and after (r = 0.680, P < 0.01). In addition, the max IMT and the hemodynamic parameters (PSV, EDV, S/D, ACCEL, TAMEAN) showed moderate agreement (ICC=0.73, 0.52, 0.58, 0.70, 0.61, 0.51, respectively). The PI, RI and HR values were excellent agreement with conventional ultrasound (ICC=0.80, 0.84, 0.94). Conclusion: About the basic assessment of carotid, the images and parameters obtained using handheld ultrasound showed a moderate to excellent agreement with conventional ultrasound. The handheld ultrasonic devices can be widely used as a diagnostic tool for carotid artery structure and hemodynamics examination.