Methods to measure protein biomarkers with ultralow detection limit (DL) and high sensitivity promise to provide valuable tools for early diagnosis of diseases such as cancer, and for monitoring therapy and post-surgical recurrence.[i, ii] Surface plasmon resonance (SPR) utilizing nanoparticle-antibody labels for signal amplification in immunoassays is an emerging approach for detecting proteins in biomedical samples.[iii–x] Herein, we show for the first time that clustering of superparamagnetic labels on SPR surfaces leads to unprecedented sensitivity and ultralow DL for protein biomarkers in serum. Specifically, antibody bioconjugates on 1 µm diameter superparamagnetic particles (MP) used for off-line antigen capture enabled SPR detection of cancer biomarker prostate specific antigen (PSA) in serum at an ultralow DL of 10 fg mL−1 (~300 aM). This approach opens doors for accurate diagnostics based on new protein biomarkers with inherently low concentrations.
No takes yet. Share an insight, caveat, or question.
Krishnan et al. (2010) studied this question.