Photometric units in Australia are currently derived from power measurements made with a roomtemperature electrical-substitution radiometer. This basis has been transferred to a cryogenic radiometer. The cryogenic radiometer was used to calibrate the absolute responsivity of a four-element, planar, transmission trap detector (measuring the transmission loss each usage), at laser wavelengths. Modelling of the quantum defect then provided a reference for absolute spectral responsivity throughout the photometric wavelength range. This was used to measure the responsivity of a number of photometers at the mercury green line and to calculate correction factors for CIE Illuminant A. The candela so derived essentially agrees with our existing unit, maintained on a set of lamps. The trap detector was also used to determine the polarization properties of the monochromator.
No takes yet. Share an insight, caveat, or question.
Gardner et al. (1998) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: