Answers are grounded in this paper and cite their sources.
The photon energy resolution of the modified SX-700 plane grating monochromator at the Finnish beamline (BL51) at MAX-laboratory in Lund, Sweden, has improved considerably after proper alignment of the monochromator. Resolving powers of about 10 000 at hν≊90 eV and 5000 at hν≊400 eV have been achieved at the first diffraction order, as inferred from total ion yield measurements at the Kr 3d5/2→5p and N2 N1s→π* resonances. The total linewidth of ≤121 meV recorded for the Ar 2p3/2→4s excitation suggests that the lifetime broadening could be smaller than 116 meV obtained from electron energy-loss spectroscopy.
No takes yet. Share an insight, caveat, or question.
Aksela et al. (1995) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: