Diffusion losses and charging efficiency were measured for three types of charge neutralizers commonly used in aerosol research: two with 85Kr and one with 210Po as radiation sources. The diffusion losses were characterized at flows of 0.5 -6 1 min−1 typically used in atmospheric aerosol physics measurements. All of the neutralizers tested exhibited high transmission efficiencies, with losses up to 25% at the smallest tested size of 3 nm, varying with size and flow in general agreement with diffusion loss theory. Charging efficiency was measured for a singly charged, monodisperse aerosol at the same flows and at concentrations of 103-104 particles cm−3. Neither of the 85Kr chargers brought the charge distribution close to equilibrium at 2 1 min−1, except at concentrations ≤ 103 cm−3. The 210Po charger produced the theoretically expected fraction of singly charged particles within the uncertainty of the experiment.
No takes yet. Share an insight, caveat, or question.
Covert et al. (1997) studied this question.