PURPOSE: The objective of this study is to directly compare different types of cervical artificial disc implants using an in vivo model capable of simulating the axial load on a neck that is similar to the human neck. METHODS: Cervical arthroplasty was performed at C3-4 in 14 healthy female adult Alpine goats. The goats were divided into three groups. Group A received Bryan (unconstrained one-piece design); Group B received ProDisc-C (semi-constrained two-piece design); and Group C received Mobi-C (unconstrained three-piece design) artificial discs. The goats were monitored in a veterinary unit for 6 months with radiography at regular intervals. RESULTS: Each goat tolerated cervical arthroplasty well and had satisfactory placement of their implant per intra-operative radiography. Implants monitored in Group A demonstrated no migration. One out of five implants in Group B experienced anterior migration at 3 months. In Group C, anterior migration and disintegration occurred in all four implants, with migration occurring during the first postoperative week in three implants and after 5 weeks in the fourth. CONCLUSIONS: Unconstrained multi-piece artificial cervical discs may be prone to anterior migration and extrusion out of the disc space. This outcome deserves attention in individuals with a hypermobile neck and/or an occupation involving the use of a head-supported mass, such as helmets.