The most perfect primary design will always leave some unwanted signal which must be coped with by the secondary unless only very high velocities are to be measured.5 The reference signal should be proportional to B and should be derived from the primary element if errors due to malphasing and to line voltage and frequency variations are to be avoided.6 Equation (3) imposes a theoretical limit on allowable resistance between electrodes, but other practical considerations suggest a threshold fluid conductivity of 20 mieromhos per cm for the present.7 With the foregoing precautions, it is realistic to expect accuracy of ±0.5 per cent of volume rate for fluids resembling tap water at velocities of as little as 1 fps, and ±1 per cent at the recommended threshold conductivity level, so far a3 magneticmeter primary elements are concerned.Some sacrifice of accuracy, of course, will be attributable to secondary instrument performance and readability.It does not appear far-fetched to say that, when handling suitable fluids, a need for wide-range accuracy alone could justify the choice of an electromagnetic flowmeter, even though the need for an obstructionless primary element were not vital.