We have constructed an ultrahigh vacuum scanning tunneling microscope (STM) for operation in the temperature range 20–300 K. The design consists of a vibration isolated sample holder mounted on a continuous flow cryostat. By rotation and linear motion of the cryostat, the sample can be positioned in front of various surface preparation and analysis instruments contained in a single vacuum chamber. A lightweight beetle-type STM head is lowered from the top onto the sample by a linear manipulator. To minimize helium convection in the cryostat, the entire vacuum system, including a liquid helium storage Dewar, can be tilted by a few degrees perpendicular to the cryostat axis, which improves the operation. The performance of the instrument is demonstrated by atomically resolved images of the Pd(111) surface and adsorbed CO molecules.