Why the study?
Is transcutaneous electrical nerve stimulation (TNS) safe for pain relief during labor without adverse effects on the mother or fetal heart?
Is transcutaneous electrical nerve stimulation (TNS) safe for pain relief during labor without adverse effects on the mother or fetal heart?
TNS with current densities not exceeding 0.5 microamperemeter/mm2 appears safe for pain relief during labor without adverse effects on the mother or fetal heart.
Answers are grounded in this paper and cite their sources.
Supports TNS safety hypothesis in labor; hypothesis-generating and should not change practice pending larger trials.
A current density standard for current shapes used in transcutaneous electrical nerve stimulation (TNS) must be established in order to avoid harmful effects. This is especially important when stimulating near vital structures such as the fetal heart. In the absence of an applicable standard, a preliminary safety norm is proposed, based on clinical experience during delivery and experimental measurements in the female bladder. Current densities due to TNS not exceeding 0.5 microamperemeter/mm2 are safe for the fetal heart. A stimulator and electrodes fulfilling the safety criteria proposed in this study have been tested. A filter which suppresses the electrical disturbances occurring during TNS, thereby permitting recording of the fetal heart rate during birth, has also been tested. TNS was given over both the low-back and suprapubic region. Results of clinical tests of the equipment during 15 supervised births are reported. No adverse effect in the mother or newborn infant were observed.
No takes yet. Share an insight, caveat, or question.
Bundsen et al. (1982) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: